Latching alarms

If the Latching Alarms selection is active, the alarm messages stay on the screen even if the
initial alarm condition goes away. This enables unattended monitoring. You will also hear a
reminder beep every 10 seconds.
To clear the message field of the no-longer active alarm messages and to clear the beep, press
Silence Alarms
the
To clear the message field of all the alarms, enabling only new upcoming alarm messages,
press the
To select latching alarms:
1.
Press the
2.
Select Install/Service and enter the password.
3.
Select Installation - Alarm Options.
4.
Select Latching Alarms - Yes.