B30 Patient Monitor
Alarm Volume Adjusts the volume of the audio alarms. The range is from 1 (soft) to 10 (loud).
NOTE: Audible alarms cannot be totally silenced with the Alarm Volume function.
Alarm Light
Adjusts the brightness of the alarm light if the light has been enabled (default) through
Monitor Setup
is from 5 to 100%. The light is on during the adjustment to help determine a suitable brightness.
Audio ON/OFF Opens a menu to select Silence Apnea, Silence ECG, Silence Apn & ECG or Silence All.
NOTE: The silencing selections that is, all other selections than the Activate Alarms) are
available only if alarm silencing has been enabled by selecting
Service - Installation - Alarm Options - Show Audio ON/OFF - Yes. By default, it is disabled. For
more details, see
Activate Alarms = Activates silenced alarms. This selection is always available for activating
alarms that have been permanently silenced using the Central.
Silence Apn = Silences apnea and disconnection alarms as well as CO
Silence ECG = Silences arrhythmia alarms and also HR limit alarms.
Silence Apn & ECG = Silences both of above.
Silence ALL = Silences permanently all alarms except FiO
Remove Menu Clears the menu selections from the display so that only 10 minute trends and limits are
displayed (push the ComWheel to return the selections on the screen).
4-6
- Install/Service - Installation - Alarm Options - Alarm Light- YES. The range
"Enabling or disabling alarm silencing" page 4-14.
Monitor Setup
- Install/
, respiration rate.
2
<18%, Ppeak >70 cmH
2
O.
2