Temperature
Measurement range:
Measurement accuracy:
Probe type:
Temperature self-check:
Time constant of temperature probes:
Reusable skin temperature probe:
Reusable adult central temperature probe:
Reusable pediatric central temperature probe:
Disposable skin temperature probe:
Disposable central temperature probe, 12F:
Disposable central temperature probe, 9F:
NIBP
Measurement range:
Typical measuring time:
Overall system accuracy:
The ESU does not cause a burn hazard through the NIBP cuff, because there is no electrical connection
between the cuff and the NIBP measuring electronics.
NOTE: NIBP measurement is intended for patients who weigh 5 kg (11 lb) and up.
NOTE: The cuff pressure measurement range is equal to cuff nominal and cuff
indication ranges.
Pulse oximetry
1
SpO
2
Automatic scaling of plethysmographic waveform.
Measurement and display range:
Calibration range:
Calibrated against functional oxygen saturation.
Measurement accuracy
Display resolution:
Wavelength of SpO
Infrared LED
Red LED
Maximum energy of SpO
Infrared LED
Red LED
Pulse rate:
Measurement and display range:
Measurement accuracy
1 The isolation barrier capacitance in the module has been minimized to reduce the hazard of burns in the event of a
defect in the ESU return electrode connection.
2 According to SP10-2002 AAMI 4.4.5.2.B, Intra-arterial method as the reference standard, mean difference of the test
system and the comparison system shall be ± 5 mmHg or less with standard deviation of 8 mmHg or less
3 Accuracy is based on deep hypoxia studies with volunteered subjects during motion and non-motion conditions over a
wide range of arterial blood oxygen saturations as compared to arterial blood CO-Oximetry. Accuracy may depend on
the sensor used, please refer to the instructions for use in the accessory package. The accuracy is expressed as rms. This
means that approximately two-thirds of the data will fall within the accuracy range.
1
10 to 45 °C (50 to 113 °F)
±0.1 °C (25 to 45 °C)
Use only GE Healthcare temperature probes or defibrillator-proof
YSI 400 series probes.
At start-up and then every 10 minutes.
adult 25 to 260 mmHg (3.3 to 34.7 kPa)
child 25 to 190 mmHg (3.3 to 25.3 kPa)
infant 15 to 140 mmHg (2.0 to 18.7 kPa)
adult less than 30 s, infant less than 25 s
Meets or exceeds SP10-2002 AAMI standards
0 to 100 %
70 to 100 %
3
:
100 to 70 %, ±2 digits
69 to 0 %, unspecified
1 digit (1% of SpO
probe LEDs:
2
940 nm
660 nm
probe LEDs:
2
42 µJ/pulse
62 µJ/pulse
30 to 250 bpm
4
:
±5 % or ±5 bpm, whichever is greater
3 s
6 s
4 s
3 to 6 s
5 to 8 s
5 to 8 s
)
2
System description
2
2-33