Points to note

You can connect and disconnect airway modules like any other modules.
For a comprehensive list of accessories, see the "Supplies and Accessories" catalog.
In the monitoring system, use only one module for measuring airway gases.
The message 'Sampling line blocked' may result if you attach the sampling line to the
water trap after turning the monitor on.
When the warning symbol is displayed beside the O
below 21%.
WARNING
Use only approved accessories. For a list of approved supplies and
accessories, see the "Supplies and Accessories" catalog delivered with the
monitor. Other accessories may cause a safety hazard, damage the
equipment or system, result in increased emissions or decreased immunity
of the equipment or system or interfere with the measurement.
CAUTION
Remove the airway sampling line from the patient's airway while nebulized
medications are being delivered.
CAUTION
Do not apply pressurized air to any outlet or tubing connected to the monitor.
Pressure may destroy sensitive elements.
WARNING
Always test the airway adapter for a tight connection and proper operation
before attaching it to the patient.