Submenu
a
Temperature
a
Service
Date&Time
(Extended Menu only)
Communication
(Extended Menu only)
Relays/Analog
(Extended Menu only)
NewSensor
(Extended Menu only)
Calibration
(Extended Menu only)
a.
These submenus can be accessed from both Menus (Main and Extended); except the Service
submenu, they contain more parameters and commands in the Extended Menu.
b.
There are four modes: Latch, Normal, Force ON and Force OFF.
Part 17997
• To read the current temperature of the Hydran 201 sensor.
• To set the parameters of the sensor temperature alarm.
• To adjust the set point of the sensor temperature.
• To set periods A and B.
• To read the actual set point, heating power and temperature of
the heating plate.
• To set the parameters of the dynamic oil sampling system.
• To read the value of the H201 sensor parameters.
• To test the H201 sensor operation.
• To read the battery voltage.
• To set the parameters of the battery alarm.
• To read the service data of the H201 sensor.
• To set the date.
• To set the time.
• To select the communication channel: RS-232 link or
supervisory link.
• To set the communication parameters.
• To read the state of the alarm relays: On or Off.
• To set the operation mode of the three alarm relays
• To select the triggering conditions of the fail alarm
(FaultTrig.).
• To read the state of the analog output and TDM signal (in %).
• To set the operation mode of the analog output and TDM signal.
• To set the H201 sensor parameters.
• To install a new H201 sensor.
• To calibrate the three analog inputs:
– The H201 sensor signal
– The H201 sensor temperature
– The heating plate temperature
• To calibrate the analog output.
Rev. 6, March 2011
Main and Extended Menus
Functions
b
.