6.4.2

Verifying and Setting Historical Data File Parameters

For details on the parameters and commands in the submenu History, see Section 4.5.1 on
page 4-12.
1.
Access the submenu H201Ti V2.2x;History.
2.
Access the submenu History;HistorySetup.
3.
Verify and set, as required, the Short Term logging rate and the four Long Term
logging hours.
4.
Access History;Clear Short Term.
5.
Press ENTER twice to erase the contents of the historic data file Short Term.
6.
Access History;Clear Long Term.
7.
Press ENTER twice to erase the contents of the historic data file Long Term.
Do not erase the contents of the historic data file Events, in order to
WARNING
CAUTION
keep a record of changes performed during the present configu-
ration.
6.4.3

Verifying and Setting Gas and Trend Alarm Parameters

For details on the parameters and commands in the submenu Gas, see Section 4.5.2 on
page 4-16. The operation of the relays and alarms is explained in detail in Chapter 10.
1.
Wait a minimum of 30 minutes after powering up the Hydran 201Ti or until the level
of dissolved gases in oil read by the H201Ti is stable.
2.
Access the submenu H201Ti V2.2x;Gas.
3.
Verify and set, as required, the 15 parameters (alarm set points and periods). Do not
set the parameters HourlyTr Period and DailyTr Period.
Part 17997
Rev. 6, March 2011
Configuring the Hydran 201Ti
6-31