Figure D-3 - Wiring Of Rs-485 Link Cable Between Two Or More H201Ci-C's Or; Table D-12 - Sw1 Voltage Selector For The Hydran 201Ci-1; Table D-13 - J10 Db-9 Male Rs-232 Connector For The Hydran 201Ci-1 - GE Hydran 201 i Instruction Manual

Hide thumbs Also See for Hydran 201 i:
Table of Contents

Advertisement

Voltage Selector
115 V
230 V
a.
The voltage selector is factory-selected and configured as per user requirements.

Table D-13 - J10 DB-9 Male RS-232 Connector for the Hydran 201Ci-1

Terminal
2
RxD (receives data)
3
TxD (transmits data)
5
GND (signal ground)
1, 4, 6, 9
DCD, DTR, DSR and RI wired together
7, 8
RTS and CTS wired together
H201C -1/H201C -C
i
Main
Terminal
Block
21
S-
22
S+
23
Figure D-3 - Wiring of RS-485 Link Cable Between Two or More H201Ci-C's or H201Ci-1's
Part 17997

Table D-12 - SW1 Voltage Selector for the Hydran 201Ci-1

a
Voltage Range
90–132 Vac, 47–63 Hz
180–264 Vac, 47–63 Hz
Name
Direction
To H201Ci-1
From H201Ci-1
i
Rev. 6, March 2011
Hydran 201Ci-1 One-Channel Controller
Comment
130-W maximum heating @ 264 Vac
Comments
The DB-9 connector is wired as in
a PC: Cross wires 2 and 3 when
connecting to a PC; do not cross
wires 2 and 3 when attaching a
modem. Only wires 2, 3 and 5 are
needed to communicate.
H201C -1/H201C -C
Main
Terminal
Block
Not
connected
i
i
21
S-
22
S+
23
D-11