Chapter 6 • Hydran 201Ti Installation
3.
Push DOWN five times until Communication appears.
4.
Access this menu by pushing ENTER.
5.
Modify the Comm Channel by pushing CHANGE.
6.
When asked to enter the password #1 (1253), use the UP and ENTER keys.
7.
Change the Comm Channel to H201Ti-DB9 using the UP and ENTER keys.
8.
Modify the Default Channel by pushing DOWN and CHANGE.
9.
Change the Default Channel to H201Ti-DB9 using the UP and ENTER keys.
10. Push DOWN four times and ENTER to modify the Gas Monitor ID.
11. Write down the current value of the Gas Monitor ID.
12. Change the Gas Monitor ID to 1 using the UP, DOWN and ENTER keys.
13. When all is done, push END.
14. Start the Hydran Host software. Please refer to the Hydran Host Software Manual.
15. In the opening window, go to the Main Menu and choose Network Setting.
16. Check [] the station No. 1 and uncheck [ ] all the other stations.
17. Return to the Main Menu.
18. Select the Real-Time Logging menu.
19. Set H201Ti List to Same as Network and set Log Rate to 5 seconds.
20. Return to the Main Menu and then to the opening window.
21. Verify that a new line of information is logged on the PC screen once every five
seconds.