Changing Max. /Min. Slice Level ([%T])
This is a command to change Max. /Min. Slice Level by specifying a binary label group A - H. The max/min. slice
level of the currently selected product type can be changed.
Command format: Comma separated
Communication Direction
From external device to PV310
Response: At Normal
Response: Error 1
Response: Error 2
ERROR signal = "ON"
Error Code: E0175 is
displayed on the screen
To specify the command format (Comma separated/Fixed Length), select "ENVIRONMENT" > "Serial
Communication" > "Command" from the menu bar.
268
Command / Description
%
T
?
?
,
a
?? : 01 to 08 (Specify the number of Slice Level in 2-digit form)
Slice Level
A
Specified number 01 02 03 04 05 06 07 08
aaa : 000 to 254 or 0 to 254 (Slice level-min.: 1 to 3 digits)
bbb : 001 to 255 or 1 to 255 (Slice level-max. 1 to 3 digits)
For example, if slice level =1, Min.=20 and Max.=255, the command is %T01, 20,
C
C
255
or %T01, 020, 255
R
R
%
T
?
?
,
a
The same command as the sending command from external device to PV310 is
transmitted.
%
U CR
Cause:
Command in the unspecified form or not provided was received.
Examples:
-
%T1,80,255 CR (any other checker number than 01 - 08 was specified)
%B01 CR (not provided command)
-
%
Z CR
Causes:
The max. value of Slice Level specified by a sending command is smaller than
the min. value of Slice Level.
The max. value of Slice Level is equal to the min. value of Slice Level.
Max./min. values of Slice Level specified by sending command are not within
the range between 000 or 0 and 255.
a
a
,
b
b
b CR
B
C
D
E
F
G
a
a
,
b
b
b CR
H