Honeywell MicroDock II Operating Manual page 61

Automatic test and calibration station
Table of Contents

Advertisement

Problem
No drive letter is created for the
MicroDock II (Windows XP).
MicroDock II Automatic Test and Calibration Station
Possible Cause
Windows XP is not correctly
mapping the drive.
Problem transferring to the
MMC card during a calibration
or bump test.
Size of MMC is not supported. Use only 128MB Delkin Multi-
M M C i s n o t c o r r e c t l y
formatted.
Log file has exceeded maximum
size.
IR not working when detector is
inserted into the module.
Table 20. Troubleshooting Tips
Solution
Ensure Windows XP assigned a
drive letter to the MicroDock II.
Solution:
1) Right-click My Computer
and then click Manage.
2 ) U n d e r C o m p u t e r
Management (Local) click
Disk Management
3) In the list of drives in the right
pane, right-click the new drive
and then click Change Drive
Letter and Path(s).
4) Click Change and in the
drop-down box, select a drive
letter for the new drive (cannot
be assigned to a mapped
network drive).
5) Click OK and then click OK
again.
Repeat the test.
Ensure the MMC card is
inserted.
media card (MMC)
Refer to
Format the MMC
Card.
Replace with a new MMC
card.
Refer to
Format the MMC
Card.
Replace with a new MMC
card.
Check for dirt on the sensor.
C o n f i r m t h e d e t e c t o r i s
activated.
Charge the detector battery if
too low.
61