Honeywell MicroDock II Operating Manual page 40

Automatic test and calibration station
Table of Contents

Advertisement

Select
Load existing configuration file from a MicroDock
Load existing configuration file from hard drive
Load default configuration file
10.
The Device Configuration dialog box displays.
11.
Select which device to configure by clicking its tab.
12.
Configure the device.
Ensure the Updatable checkbox is checked.
13.
Select a program option that will reprogram the detector. Each detector's
configuration screen has different reprogram options. Refer to Configuring
Detectors in the Fleet Manager II User Manual.
14.
When the configuration is complete, click
dialog box displays.
Select Save to MicroDock(s)
15.
The Save to MicroDock(s) dialog box displays.
16.
Select a MicroDock II to save to and click
17.
A dialog box displays the message S
Configuration File. Click OK.
The detector configuration will be updated when the detector is inserted into
the docking module, and a bump test or calibration is initiated.
18.
To reconfigure the detector, press
If Reprogram with user's consent was selected under Program options,
Reconfigure detector? displays on the base station LCD.
Press
calibration.
Press
calibration.
40
Yes to reconfigure the detector and proceed to the bump test or
No to bypass reconfiguration and proceed to the bump test or
Save. The Save Configuration
OK.
uccessfully saved MicroDock
BUMP CHECK or
MicroDock II Automatic Test and Calibration Station
CALIBRATION.