Calibration
Guidelines
CAUTION
If adjustments are made to the real-time clock, the calibration due date will be
invalid for the XD and X5.
Recalibrate the detector immediately.
When calibrating the sensors, adhere to the following guidelines:
•
Recommended gas mixture:
O
2
CO
CO
H
S (high range)
2
H
S (low methanol)
2
H
S
2
PH
3
SO
2
NH
3
NH
(high range)
3
NO
2
HCN
NO
Cl
2
VOC
CO
2
LEL
•
Calibrate only in a clean atmosphere that is free of background gas. Do not
operate the station in a hazardous area. Failure to adhere to this guideline
can result in possible personal injury and/or property damage.
•
Honeywell Analytics recommends using premium grade calibration gases
and cylinders that are certified to National Standards. The calibration gases
must meet the accuracy of the detector.
•
Do not use a gas cylinder beyond its expiration date.
•
All calibration cylinders must be used with a demand flow regulator and
must meet the following maximum inlet pressure specifications:
i.
Disposable cylinders 0-1000 psig/70 bar
ii.
Refillable cylinders 0-3000 psig/207 bar
•
Refer to the detector user manual for recommended calibration
frequencies.
MicroDock II Automatic Test and Calibration Station
clean air, 20.9% vol.
(low H
sensitivity): 50 to 500 ppm balance N
2
50 to 500 ppm balance N
10 to 100 ppm balance N
10 to 100 ppm balance N
10 to 100 ppm balance N
1 to 5 ppm balance N
10 to 50 ppm balance N
20 to 100 ppm balance N
20 to 100 ppm balance N
5 to 50 ppm balance N
5 to 20 ppm balance N
10 to 250 ppm balance N
3 to 25 ppm (for bump checks only)
100 ppm isobutylene
5000 ppm balance air or 0-5.0 % v/v CO
10 to 100% LEL or 0.5 to 5% by vol. methane
balance air
2
2
2
2
2
2
2
2
2
2
2
2
2
51