Inserting the XD

CAUTION
Infrared or intense ambient light (sun or halogen) may interfere with the station/
detector communication.
1.
Activate the detector and wait until it is in normal operating mode.
2.
Ensure the alligator clip is closed and the ring is resting flat to prevent
disruptions with the transmission.
3.
Press the two release tabs on the docking module and raise the lid.
4.
Lower the detector (LCD face down) into the detector bay. Push forward to
ensure the top of the detector connects with the top of the bay.
5.
Lower the lid and press down until the release tabs click.
When the detector has been inserted correctly, the RUN LED(s) on the docking
module lights yellow and Unit Inserted displays.
The station LCD displays the following docking module identification; the
typ
e of detector that is inserted, and
bay and serial number of the docking module.
6.
To bypass the
station.
MicroDock II Automatic Test and Calibration Station
Figure 5. Inserting the XD
Item Description
1
Alligator clip
2
Docking module lid
3
Release tabs
4
Detector bay
Table 7. Inserting the XD
station identification screen, press any
button on the base
17