Sensor Disabled

WArNINg
To prevent personal injury and/or property damage, replace the sensor
immediately. Refer to the corresponding detector user manual for sensor
replacement procedures.
If a sensor is disabled, results from the sensor do not display on the station LCD.
– displays where the gas type typically displays on the LCD.
NOTe
If a sensor is disabled, it will not be recorded in the results log.

Data Transfer

(XD only)

Transferring Datalogs

The data transfer function is used to transfer datalogs from the detector to the
station's MMC card.
Depending upon the amount of information stored in the detector, the transfer of data
may require 12-14 minutes to complete. To transfer a datalogs from the detector to
the docking module and then to a MMC card, complete the following:
1.
Activate the detector and wait until it is in normal operating mode. Insert it
into the docking module.
2.
Activate the station.
3.
Press
flashes rapidly and the docking module begins to transfer the datalogs to
the station.
When the transfer process is complete, the docking module indicates the
status of the transfer as PASS or FAIL.
Successful Transfer: If the datalog transfer is successful, the PASS LED
lights green, the RUN LED flashes yellow slowly, and the detector beeps.
Unsuccessful Transfer: If the datalog transfer is unsuccessful, the FAIL
LED lights red and the RUN LED flashes slowly.
If a failure occurs during the transfer of a datalog, typically some of the data
transfers to the MMC card. To determine if any data has been transferred,
access the MMC data using Fleet Manager II.
If required, repeat step #3 to transfer the datalog.
After a successful transfer to the station, the station transfers the datalog to
the MMC. The RUN and the PASS LEDs light on the docking module and
the following screen displays on the station.
MicroDock II Automatic Test and Calibration Station
DATA TRANSFER on the docking module. The yellow RUN LED
55