Results History

The results history function is used to display the results of the last ten records of
both bump checks and calibrations (combined total).
The results display from most recent to the oldest record. Test errors display as the
type of error; Unit removed, MMC error, and IR error.
NOTe
The results history log is erased when the station is deactivated.
To view the results history log, complete the following:
1.
From the normal operating screen, press
The most recent history log (1) displays.
2.
To access the next log, press
remaining logs.
3.
Press

Reconfiguring the Detector

Fleet Manager II can reconfigure XD and X5 detectors with the base station. The
detector can be reconfigured prior to performing a bump test or calibration. To
reconfigure a detector, complete the following:
1.
Activate the base station. Ensure it is in normal operation.
2.
Ensure that a correctly formatted MMC card is inserted in the base station.
See
Inserting/Replacing a MMC
3.
Connect the USB cable to the computer and to the USB port on the base
station.
4.
Activate the detector and wait until it is in normal operation.
5.
Insert the detector into the detector bay.
6.
From the computer, open Fleet Manager II.
7.
Login to the Administration functions.
8.
Under the Devices tab, click Configure Devices.
9.
The Load Configuration File dialog box displays.
MicroDock II Automatic Test and Calibration Station
OK to return to the normal operating screen.
log.
^ . Continue to press
Card.
^ to view the
39