Accessing the MMC Card During a Test

If an attempt is made to access the MMC card on the station from a PC while a bump
check or calibration is being performed, the following screen displays.
NOTe
Honeywell Analytics recommends that bump checks and calibrations be
completed before attempting to access test results from the PC.

Order of Gases Applied for Bump Checks and Calibrations

WArNINg
It is extremely important that gases are applied in the order that is specified in
this manual.
Failure to adhere to the rules in Table 17 and Table 18 of this manual will
result in incorrect sensor readings and possible personal injury and/or property
damage.
NOTe
Manual application of gas types is available for the X5 detectors only.
To manually apply gases for bump checks and calibrations, complete the
following:
1.
Review
Application
2.
Determine the gases that are to be applied.
3.
Using
Table
calibrated on the top row.
4.
Within the far left column, locate the next gas type (sensor) to be bumped
or calibrated and refer to the rule number (if applicable).
5.
Using
Table
desired order to apply gases corresponds to the gas application rules.
6.
Continue to reference the table and rules to ensure the gases will be applied
in the correct order.
7.
When the correct order is determined, proceed to the
Calibration
MicroDock II Automatic Test and Calibration Station
Table 17. Gas Type Application
Rules.
17, locate the first gas type (sensor) to be bump checked or
18, locate the corresponding rule number. Ensure that the
section.
Table, and
Table 18. Gas
Bump Check
or
45