6.
When the required inlet value displays, press
and deactivate the field. The * no longer displays to the right of the inlet
indicating the conflict has been resolved.
Unsuccessful Conflict Resolution: If the conflict is not resolved by selecting a
different inlet, refer to the following:
Ensure that the selected inlet is correct.
Ensure the gas cylinder that is attached to the selected inlet is correct.
Press

Not Found

If a value of 1 or more displays before Not Found, complete the following:
Or
Press
cancel to quit the current bump check/calibration.
1.
Press
cannot be found.
The * icon displays on the far right side on the row of the gas that cannot
be found. The station automatically defaults to the Purge inlet.
2.
Press
Press sel to select a different inlet.
NOTe
For gases not found: If sel is not selected within 30 seconds, the station
automatically defaults to the Purge inlet.
3.
If required, repeat steps #1-2 for any additional gases not found and confirm
the following:
Ensure that the selected inlet is correct.
Ensure that the gas cylinder attached to the selected inlet is correct.
Press
44
OK to accept the station's default selection.
more to access the gas type/inlet screen to view which gas type(s)
OK to accept the default Purge inlet.
OK to accept the station's default selection.
MicroDock II Automatic Test and Calibration Station
sel to confirm the selection