Setup/Configure the Monitor

Set Patient-Monitor Type

3-10
Installation: Setup/Configure the Monitor
Setting up the monitor requires you to configure the monitor at the
Service mode level and at the Boot Loader level.
Choose and program the procedures listed below in the order presented.
Each procedure is described on the following pages.
Set Patient-Monitor Type (Service mode)
Set Monitor Defaults Password (Service mode)
Set Unit Name (Service mode)
Set Bed Number (Service mode)
Set Country Selection (Boot Loader)
Set Language (Boot Loader)
127(
Refer to the operator's manual to configure this monitor for clinical
use.
The Patient-Monitor Type selection determines the type of monitor
desired, i.e adult, neonatal or operating room. Different alarms and
parameters are activated for each selection. This menu item is part of the
SERVICE MODE menu.
Access Patient-Monitor Type option, starting from the Main Menu.
1. Select MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the Trim Knob control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407)
3. Select PATIENT-MONITOR TYPE. Be sure to read the information
in the Attention box before changing anything.
4. Rotate Trim Knob control to select the type of environment the
monitor will be used in.
5. Press Trim Knob control to exit. Your selection displays at the top of
the screen after the time.
Transport Pro Patient Monitor
2012659-002
042A
Revision A