1.4 Analyzer Modules
The Vertex M System is populated with one or more
types of analyzer modules. Each system may contain
Chemcassette
modules or Pyrolyzer modules.
®
Modules are installed in slots.
CC
Tier 1
Slot 1
Table 1-1. Module Tier Structure
Chemcassette
modules occupy one slot each.
®
However, Pyrolyzer Chemcassette
installed in slots 1 and 2. Pyrolyzer configuration and
status information will appear in slot 2.
Number
of Points
Chemcassette
®
8
Pyrolyzer
8
Table 1-2. Required Slots
Examples of possible combinations in a Vertex M
System:
•
One, Two or Three Chemcassette
•
One Pyrolyzer Chemcassette
•
One pyrolyzer Chemcassette
Chemcassette
®
Your monitor will include only those modules specified
at time of ordering.
Section 1 - Introduction
CC
PYRO
Slot 2
Slot 3
modules are to be
®
Installed
Possible
into Slots
per Vertex
M System
1, 2 or 3
3
1 and 2 Only 1
modules
®
module
®
module, one
®
module
1.5 Sampling System
Each Analyzer module is a monitoring center for
sampling lines from sample locations. As they apply to
the Vertex M System, the words point, line and location
require definition:
•
A location is a place to be monitored
CC
•
Sample atmosphere runs from the location to
the Vertex M System via a line
•
Each of the 24 sample tubing connections on
the Vertex M System corresponds to a point.
A sample line can be connected directly to
a single point or multiple points via a 4-port
manifold
The system draws air simultaneously from all locations.
Total
Two different types of flow are:
•
Transport flow: high-velocity, large-volume air
movement through the lines
•
Sample flow: air admitted to the
Chemcassette
The high speed of transpor t flow allows rapid
monitoring and response time when using long lines
from monitored locations to the Vertex M System. A
small portion of the transport flow (sample flow) is
analyzed to determine concentration levels.
The complete sampling and monitoring system consists
of the following components:
•
Sample lines to all monitored locations
•
Flow connections through quick-connect ports
in bulkheads on top of unit
•
Moving cable and connectors
•
Vacuum pumps
•
Analyzers incorporating manifolds,
Chemcassette
•
Flow controlling proportional valve
1-13
detection system
®
and filters
®