Analyzer Modules; Sampling System - Honeywell Vertex M Technical Handbook

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Vertex M
24-Point Continuous Monitor
TM

1.4 Analyzer Modules

The Vertex M System is populated with one or more
types of analyzer modules. Each system may contain
Chemcassette
modules or Pyrolyzer modules.
®
Modules are installed in slots.
CC
Tier 1
Slot 1
Table 1-1. Module Tier Structure
Chemcassette
modules occupy one slot each.
®
However, Pyrolyzer Chemcassette
installed in slots 1 and 2. Pyrolyzer configuration and
status information will appear in slot 2.
Number
of Points
Chemcassette
®
8
Pyrolyzer
8
Table 1-2. Required Slots
Examples of possible combinations in a Vertex M
System:
One, Two or Three Chemcassette
One Pyrolyzer Chemcassette
One pyrolyzer Chemcassette
Chemcassette
®
Your monitor will include only those modules specified
at time of ordering.
Section 1 - Introduction
CC
PYRO
Slot 2
Slot 3
modules are to be
®
Installed
Possible
into Slots
per Vertex
M System
1, 2 or 3
3
1 and 2 Only 1
modules
®
module
®
module, one
®
module

1.5 Sampling System

Each Analyzer module is a monitoring center for
sampling lines from sample locations. As they apply to
the Vertex M System, the words point, line and location
require definition:
A location is a place to be monitored
CC
Sample atmosphere runs from the location to
the Vertex M System via a line
Each of the 24 sample tubing connections on
the Vertex M System corresponds to a point.
A sample line can be connected directly to
a single point or multiple points via a 4-port
manifold
The system draws air simultaneously from all locations.
Total
Two different types of flow are:
Transport flow: high-velocity, large-volume air
movement through the lines
Sample flow: air admitted to the
Chemcassette
The high speed of transpor t flow allows rapid
monitoring and response time when using long lines
from monitored locations to the Vertex M System. A
small portion of the transport flow (sample flow) is
analyzed to determine concentration levels.
The complete sampling and monitoring system consists
of the following components:
Sample lines to all monitored locations
Flow connections through quick-connect ports
in bulkheads on top of unit
Moving cable and connectors
Vacuum pumps
Analyzers incorporating manifolds,
Chemcassette
Flow controlling proportional valve
1-13