Vertex M
24-Point Continuous Monitor
TM
3.6.3 Set Analyzer Window
To configure the type of analyzer slot, press the top of the button representing the slot. The Set Analyzer Window
opens.
Analyzer Type
Choose the model of analyzer populating
the slot. A pyrolyzer takes up slots 1 and 2.
If pyrolyzer is selected, slot 1 automatically
becomes unavailable.
Any analyzer that is physically installed but
configured on this screen to "None" should
be de-energized.
Family/Chemcassette
Choose the family of target gases.
Target gases for all eight points in
the analyzer must be in the same
ChemCam AutoPicture
When either ChemCam for Level 1 or Level 2 gas is selected, a gas alarm signals the Vertex M System to
store a picture of the stain the next time it advances the Chemcassette
tape stops with the stain under the camera, the ChemCam takes a picture and then advance continues. The
ChemCam field of view is only wide enough to capture four points on one picture. See
ChemCam
for additional information on ChemCam features and functions.
Note:
This activity will consume additional tape. Some faint stains may not be visible via the camera
Section 3 - Startup
family of gases.
If the optional ChemCams are installed, select under what condition
a picture of the Chemcassette
Options
When you choose a family of gases, Vertex M
enters default settings into the profile. Selecting
"Override Defaults" allows you to change the
target gas type, alarm levels and the location
assignment.
ChemCam AutoPicture
®
stain will be automatically taken.
®
3-20
tape. When the tape advances, the
Section 4.5.3 Optional