6.5 Information Events

The Vertex M System enters informational and other non-fault events into the event history database. These
do not require any action by the user.
Use the event history to check the status of the monitor.
Event Type In History
ALM SIMULATE
Various
ALM 1
Concentration > AL1
ALM 2
Concentration > AL2
ALM WARNING
Concentration > Warning
AZ NO RECORD
Various
AZ PROGRAM INSTALL Various
CALIBRATION
Command – Valves Auto Balance
CONFIG INSTALL
Various
DEFAULT RESET
Various
FAULT SIMULATE
Various
INFO
20 mA Loop Test Ended
INFO
20 mA Loop Driven for Test
INFO
Window Zero Reset
INFO
Analyzer accepts new location
INFO
Analyzer Powered Up
INFO
Az Button Resets Alm&Flts
INFO
Az Button Starts Monitoring
INFO
Az Button Stops Monitoring
INFO
Clock Adjusted By DAq PC
INFO
Flow Auto Balance Performed
INFO
Hardware Configuration Changed
Section 6 - Troubleshooting
Event History Message
An alarm was artificially created. See
Gas concentration exceeded limit
Gas concentration exceeded limit
Gas concentration exceeded lower detection limit
A fault was cleared in the DAq for which the analyzer has no record
The analyzer program update process was started
Auto Balance was started
The configuration profile installation was started
A fault was cleared in the DAq from an analyzer that is no longer present in the rack
A fault was artificially created. See
User initiated action
User initiated action
Optics defect detected and corrected
Analyzer moved
User initiated action
Internal override pushbutton operated
Internal override pushbutton operated
Internal override pushbutton operated
Normal drift or clock set on DAq
User initiated action
Auto Balance, Load CC, New Location or Pump Alternated
6-17
Possible Cause
Section 4.6.6 Security
Section 4.6.6 Security