2.4.2 Sample Line Connections

Tubing
Stop
Figure 2-1. Sample Line Inlet Connections
To prepare for installation of sample lines, remove the
FEP Teflon tubing from the installation kit. The top of
the unit includes 25 connections:
24 Sample Inlets
Exhaust Outlet (See
Pump Exhaust Line
Note:
Always perform a leak check after installing sample
lines. See
Section 3.8 Leak Checking Sample Lines
for the leak check procedure.
Each inlet has a quick connect/disconnect fitting
with an internal O-ring and an external grab ring. To
Section 2 - Installation
Correct - Fully Inserted
Incorrect - Not Fully Inserted
O-Ring
Grey Locking Insert
Section 2.5 Installing
for connection.)
install a tube into a sample line inlet, insert the tube
far enough into the fitting to ensure that the tube has
passed through both the external grab ring and the
internal O-ring and is firmly seated against the stop. The
insertion depth for a correctly installed sampling line is
1/2 in. to 5/8 in. (12 mm–16 mm). Verify the insertion
depth by holding the tube and marking with your thumb
where it emerges from the fitting. Remove the tube to
measure the insertion depth.
CAUTION
Improper installation of the tube into the con-
nector results in dilution of the sample.
2.4.3 Installing Sample Line Particulate
Filters
Attach a sample line filter to the sampling end of the
line for all locations.
CAUTION
Keep in mind that excess amounts of dirt in
the filters reduces the sample flow, raises
sample vacuum and may affect concentration
readings of the analyzer.
See
Appendix A Specifications
filter type to use with each target gas.

2.5 Installing Pump Exhaust Line

This section describes exhaust connections and
installation. The Vertex M System is equipped with a
vacuum pump that is located in the bottom of the Vertex
M System cabinet. The pump exhaust line connects
to the manufacturing facility central toxic exhaust
system.