Figure 3-30: Mounting In Concrete Silos - Honeywell SLG 700 User Manual

Smartline level transmitter guided wave radar
Hide thumbs Also See for SLG 700:
Table of Contents

Advertisement

Figure 3-30: Mounting in concrete silos

Remote mount
3.2.10.1
In applications where a remotely mounted display is required, the remote mount allows
the electronics housing to be mounted 3m away from the process connector. This can be
useful when access to the mounting location is limited. To assemble the remote mount,
attach the process connector first, followed by securing the mounting bracket to a pipe or
wall. Secure the electronics module to the bracket with the 3 supplied M6 screws.
Connect the cable and check bends for minimum radius (see
Figure
3-31) to prevent
damage. Torque the 2 nuts to 6Nm (4.4ft-lbs).
SLG 700 SmartLine Level Transmitter User's Manual
Revision 7