Figure 3-24: Flange Mounting; Table 3-16: Slg720: Recommended Nozzle Dimensions - Honeywell SLG 700 User Manual

Smartline level transmitter guided wave radar
Hide thumbs Also See for SLG 700:
Table of Contents

Advertisement

Nozzle mount
3.2.9.2
The transmitter can be mounted to a tank nozzle using the appropriate flange.
shows recommended nozzle dimensions based on probe type.
Recommended nozzle diameter (D)
Minimum nozzle diameter (D)
Recommended nozzle height (H)
(*) When using a flexible probe in nozzles taller than 6" (150mm) the SWB wire probe with
extension stud is recommended. SWB is an option in the model selection guide. It offers a
300mm rod extension to keep the selection of the wire probe that is in the nozzle, from moving.
In certain applications taller nozzles may be accommodated but near zone performance at the
exit of the nozzle may be reduced. For nozzle dimensions that do not meet the requirements
outlined in
Table 3-16
Page 64

Figure 3-24: Flange mounting

Table 3-16: SLG720: Recommended nozzle dimensions

Single probe
6" (150mm)
2" (50mm)
4" (100mm) + nozzle
diameter (*)
contact the Honeywell Technical Assistance Center.
SLG 700 SmartLine Level Transmitter User's Manual
(rod/wire)
> probe diameter
> probe diameter
N/A
Table 3-16
Coaxial probe
Revision 7