Mounting on a bypass / bridle
3.2.9.5
SLG 700 transmitter can be successfully installed in a new or existing bypass pipe,
bridle, or a side pipe as shown in
and allows the addition of radar level measurement to an otherwise busy installation.
A similar installation is also possible inside the main container, when installing the SLG
700 transmitter on a stilling well.
N = Inlet diameter
L = Effective measurement range (≥ 12"/300mm)
D = Bypass diameter (N<D)
Table 3-17: SLG720 bypass and stillwell recommended diameters
Probe type
Rod probe
Wire probe
Coaxial probe
Revision 7
Figure
Figure 3-28: Bypass installation
Recommended diameter
3" or 4" (75mm or 100mm)
4" (100mm)
N/A
SLG 700 SmartLine Level Transmitter User's Manual
3-28. This type of installation is often simpler
Minimum diameter
2" (50mm)
2" (50mm)
1.5" (37.5mm)
Page 67