<Return>
Trim Zero
DAC Trim
(HART only)
Trim Span
Note: Loop must
be removed from
Automatic
Control
Set DAC Normal
<Return>
Set DAC Output
Loop Test
(HART only)
Note: Loop must
be removed from
Set DAC Normal
Automatic
Control
<Return>
Poll Address
Final Assy Num ###
Show Date
HART Params
(HART only)
Year
Month
Day
Write Date
Page 98
This selection will calibrate the loop zero
output to 4.000 mA.
Connect a current meter to the transmitter to
monitor the loop output.
Note: The TEST +/- terminals should not be
used to trim the transmitter.
Press Enter to set the loop output to 4mA.
When the prompt "Enter reading" appears,
enter the value shown on the current meter (in
milliamps) and press Enter again. The
transmitter adjusts the DAC output to 4mA.
This selection will calibrate the loop span
output to 20.000mA
Connect a current meter to the transmitter to
monitor the loop output.
Note: The TEST +/- terminals should not be
used to trim the transmitter.
Press Enter to set the loop output to 20mA.
When the prompt "Enter reading" appears,
enter the value shown on the current meter (in
milliamps) and press Enter again. The
transmitter will adjust the DAC output to 20mA.
This selection allows the loop to be returned to
its Normal mode (Automatic Control) after
performing the Trim operation.
This selection allows you to force the DAC
output to any value between 3.8 and 20.8mA.
Note: This selection will put the DAC into
Fixed Output Mode.
This selection allows the loop to be returned to
its Normal mode (Automatic Control) after
performing the Trim operation.
##
Yes
No
# # # #
January -
December
# #
Press Enter to write the HART Date to the transmitter.
SLG 700 SmartLine Level Transmitter User's Manual
0 (default) to 63
Asset tracking number
If Yes selected HART Date
Options will be visible and
can be configurable
Enter the current year.
Select the current month.
Enter the day of the month.
Press  to enter menu
selection.  and  to
select number.
 to enter and shift to
next digit
Press  to enter menu
selection. Scroll to Set
DAC Normal
Press  to initiate
Press  to enter menu
selection
Scroll to Set DAC Normal
Press Enter to initiate
 and  to select number.
 to enter and shift to
next digit
Press  to enter menu
selection. Scroll to Set
DAC Normal
Press  to initiate
Press  to enter menu
selection.  and  to
select number.
 to enter and shift to
next digit
Revision 7