Depending on probe length there may be 0 to 2 coaxial outer tube "extension"
segments and 1 coaxial tube "end" segment. Slip the next coaxial tube
"extension" segment over the last rod segment and spacer and tighten to the
coaxial coupler.
6
Note: Tighten connection point to the following torque:
SLG726 30Nm (22ft-lbs)
It is recommended that a process compatible thread locking compound (i.e.
Loctite 242) be used on the outer conductor threaded joints.
Repeat steps 3-6 above until only the rod "end" segment and coax tube "end"
7
segment remain.
Attach the last "star shaped" spacer over the previous rod segment end, as in
8
step 3 above.
9
Attach rod "end" segment as in step 4 above.
Slip the coaxial coupler over the rod and spacer and tighten.
Note: Tighten connection point to the following torque:
10
SLG726 30Nm (22ft-lbs)
It is recommended that a process compatible thread locking compound (i.e.
Loctite 242) be used on the outer conductor threaded joints.
Slip the coaxial end segment over the rod end segment and secure to coaxial
coupler.
Note: Tighten connection point to the following torque:
11
SLG726 30Nm (22ft-lbs)
It is recommended that a process compatible thread locking compound (i.e.
Loctite 242) be used on the outer conductor threaded joints.
Install the end spacer between central conductor and outer tube in counter-
12
bore. Secure end spacer using the retaining ring.
Insert 2 M3 set screws into each coupler.
Note: Tighten M3 set screws to 1.0Nm (8.8in-lb).
13
It is recommended that a process compatible thread locking compound (i.e.
Loctite 242) be used on each set screw.