Figure 2-10 Power Wiring - Models With Ce Mark - Honeywell DR4300 Product Manual

Circular chart recorder
Hide thumbs Also See for DR4300:
Table of Contents

Advertisement

DR4300 Circular Chart Recorder
It is the user's responsibility to provide a switch and
1
non-time delay (North America), quick-acting, high
breaking capacity, Type F, (Europe) 1/2 A, 250 V
fuse(s) or circuit-breaker as part of the installation.
2
PROTECTIVE EARTH (Ground): Grounding of
this recorder shall be in accordance with National
and Local electrical codes.
3
NOISELESS (Clean) EARTH (Ground): To minimize
electrical noise and transients that may adversely
affect the recorder, connection of the recorder terminal
(located in the rear of the case) to a local earth
(ground), using a No. 11 (4 mm
conductor, is required.
30
Main PCA
Pen 2
TB1
L N
2
) braided copper
Figure 2-10 Power Wiring – Models With CE Mark
DR4300 Circular Chart Recorder Product Manual
Main PCA
Pen 1
TB2
TB1
L
N
2
1
L1/H
100 Vac to 240 Vac
TB2
24820
Ground
Screw
3
2