Honeywell DR4300 Product Manual page 164

Circular chart recorder
Hide thumbs Also See for DR4300:
Table of Contents

Advertisement

DR4300 Circular Chart Recorder
Step
Description
2
Calibrate 0 %
3
Calibrate 100 %
4
Exit the Calibration Mode
152
Press
FUNC
FUNC
FUNC
DR4300 Circular Chart Recorder Product Manual
Action/Result
You will see:
Upper Display
APLY
Lower Display
INZERO
• Adjust your calibration device to an output signal equal
to the 0 % range value for your particular input type.
See Table 5-1 for Voltage or Resistance equivalents.
• Wait 60 seconds, then go to the next step.
You will see:
Upper Display
APLY
Lower Display
INSPAN
Adjust your calibration device to an output signal equal to
the 100 % range value for your particular input type (see
Table 5-1 for voltage or resistance equivalents). Wait 60
seconds, then go to next step.
The recorder stores the calibration constants.
The value of input set up group parameter "LD CAL" is set
to "FLD" (field).
The recorder begins to use the new field calibration values
for this channel.
Repeat this procedure for the other pen, if required.
4/00