DR4300 Circular Chart Recorder
Step
To fine-tune the alignment of the pen, press the raise switch
9
SW4 or lower switch SW3 until the pen tip is aligned with the
100 % line on the chart.
Approximately 10 seconds after the last press of the raise or
lower switch the pen will move downscale as described in
Step 2.
To exit the pen alignment cycle place SW5 to the run
10
position (toward the bottom of the board). Pen will return to
normal operation.
170
Action
DR4300 Circular Chart Recorder Product Manual
Graphic
See Figure 3-1 for the location of SW3
and SW4.
4/00