Moving To A Specific Data Point (Dr-Properties-Record) - GE DMS Go Operator's Manual

Table of Contents

Advertisement

Chapter 6. Working with Recorded Data
6.3
Using Existing Data Recorder Files (DR - PROPERTIES) (cont.)
Once thickness measurements are stored in data files, users can quickly locate specific points within a file and locate specific values for all
file parameters. For instance, users can pinpoint a value in a specific grid location, check maximum and minimum values, and t. This section
describes ways you can use the Properties submenu to examine and edit your files.
6.3.1

Moving to a Specific Data Point (DR-PROPERTIES-RECORD)

Note: To use this function, you must have created a 2-dimension file having ROWs and COLumns. A LINEAR file has LOCATIONS only!
Note: The described features are available in VIEW mode LIST only.
In the RECORD submenu, you can directly access the specific row and column location of the currently active file.
Be sure the RECORD submenu is highlighted in the PROPERTIES function (see Figure 29 on page 110).
1.
2.
With the joystick (
Move down to the COLUMN entry. Use the joystick (
3.
A red outline highlights the selected value at that point in the file. The following entry (Value) also displays the selected value, and the entry
Attachment Type describes any attachment inserted at that point. If you have selected a file containing A-scans, pushing the joystick
horizontally displays the stored A-scan.
112
Your Complete Source for
Testing Equipment. Since 1969!
) , move to the ROW entry. Use the joystick (
www.BergEng.com
Berg Engineering & Sales Company, Inc.
) to scroll to the desired selection.
) to scroll to the desired selection.
DMS Go Operator's Manual
1-847-577-3980
Info@BergEng.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

022-510-240022-510-176

Table of Contents