Troubleshooting
Repeat the scenario and compare that to the operation as described in the manual. If the
operator deviated from the manual, repeat the task using the instructions as written.
If these steps do not resolve the problem, refer to the following section for specific problems and
solutions. If the problem still cannot be resolved, contact GE Healthcare Technical Support.

Equipment Problems

This section lists the problems that are likely to occur and their solutions.

System Does Not Power Up

Verify the unit is turned on.
l
If it is not, turn on the unit. See
l
Verify the battery is installed and charged.
l
See
"System Errors" on page
and charged.
See
"Replacing the Battery" on page
Verify the unit is connected to an AC power outlet.
l
Verify the equipment is receiving power from the AC power outlet.
l
If the unit is receiving power, the Power LED will be lit.

Acquired ECG Data Displays Unacceptable Noise

Check the patient's position.
l
The patient should remain motionless during the acquisition of a resting ECG.
Use the Hookup Advisor indicator to help determine the cause of the noise.
l
For more information, see
Verify the electrodes are placed properly.
l
See
"Standard 12 Lead Placement" on page
Verify the electrodes have been applied correctly.
l
Perspiration, excessive hair, lotions, and dead skin cells must be removed from the electrode site.
See
"Preparing the Patient" on page
Check for defective or expired electrodes.
l
136
"Turning on the System" on page
138, for instructions on verifying whether the battery is installed
132for instructions on installing the battery.
"Hookup Advisor" on page
57.
55.
MAC™ 600
51.
68.
2047426-001K