Chapter 2. Instrument Setup

2.4 Installing a Probe

Follow the instructions in this section to install a probe on your DMS Go.

2.4.1 Connecting the Probe

When connecting a probe to the instrument, the following steps must be taken:
Properly complete the physical connection of the probe to the instrument.
Properly configure the instrument to work with the connected probe.
The DMS Go accepts either a single-element probe or a dual-element probe.
A single element probe has only one ultrasound crystal. It acts first like a
loudspeaker, converting the electronic pulse into a mechanical ultrasound pulse, then
the same crystal acts at the same time like a microphone, receiving all the echoes,
converting these back into electronic pulses.
54
Your Complete Source for
Testing Equipment. Since 1969!
www.BergEng.com
Berg Engineering & Sales Company, Inc.
Receive
Probe
Transmit
Probe
Figure 19: Probe Connector Locations
DMS Go Operator's Manual
1-847-577-3980