2.5.2 2-Point Calibration (PROBE & CAL - CALIBRATION)
For two-point calibration
a. The DMS Go displays "Press ENTER to edit low calibration standard thickness or couple to low calibration thickness". The
thickness of the cal block is displayed (instead of measured thickness). If you do not need to edit the value, couple the probe to the
low-calibration thickness calibration block The DMS Go indicates that it is acquiring the value, and then directs you to remove the
probe.
b. The DMS Go displays "Press ENTER to edit high calibration standard thickness or couple to high calibration thickness". The
thickness of the cal block is displayed (instead of measured thickness). If you do not need to edit the value, couple the probe to the
high-calibration thickness calibration block. The DMS Go indicates that it is acquiring the value, and then directs you to remove the
probe. Then press DONE.
6.
After completing the procedure, use the joystick (
66
Your Complete Source for
Testing Equipment. Since 1969!
) to move to another function, or press HOME to return to the Main Menu.