Honeywell MultiRAE Series User Manual page 96

Portable multi-gas monitor
Table of Contents

Advertisement

Parameters Configured by Safety Suite Device Configurator (SSDC)
Default
Parameter
Value
Man-
down
30
alarm
seconds
sensitivity
Falldown
30
sensitivity
seconds
Falldown
45
Warning
seconds
After
Man-
Down
Medium
alarm
sensitivity
Falldown
Medium
Sensitivity
* If the Falldown Time value is set to "0," the Man Down algorithm focuses only on motionless
behavior. If the Falldown Time value is not set to zero, the algorithm uses acceleration to trigger
its function.
If acceleration is sensed, and motion does not resume, then the Warning Time starts. During
this period, the instrument waits for motion. If no motion occurs by the end of the Warning
Time, the Window period is entered. During that time, the display shows the "Are You OK?"
message. The buzzer sounds and LEDs flash continuously, and a countdown begins.
If the MultiRAE's user presses [Y/+] for "Yes" in response to the "Are You OK?" question
l
on the screen before the countdown reaches zero, the Man Down alarm stops and the
main reading screen is displayed.
If the person does not press [Y/+] for "Yes" in response to the "Are You OK?" question on
l
the screen before the countdown reaches zero, the Man Down alarm is sounded and
LEDs flash continuously.
If the person presses [MODE] for "No" during the countdown, the Man Down alarm
l
starts.
Honeywell MultiRAE Series™
Range
30 to 90 seconds
30 to 90 seconds
0* to approximately 1,000 seconds.
Setting to 0 disables the falldown
function. Settings are 31 to 1000
and Falldown Time must be greater
than the Motionless Time setting.
Low (7%)
Medium (37%)
High (63%)
Custom
Low (33%)
Medium (67%),
High (100%)
Custom
91
Sensitivity
Low/Medium/High/Custom
percentage
Low/Medium/High
Low/Medium/High/Custom
percentage
Low/Medium/High
User Manual