Enabling 3-Point Calibration for MultiRAE and MultiRAE Pro Via Safety Suite
Device Configurator (SSDC)
The MultiRAE must be connected to a PC through the supplied Desktop Cradle, Travel Charger,
or AutoRAE 2 and must be in the PC or AutoRAE 2 communications mode.
1.
Start up the Safety Suite Device Configurator (SSDC) software, enter a password, and
detect the instrument following the directions provided in the SSDC User's Manual.
2.
Click on the serial number of the MultiRAE to access its information.
3.
Click on the "Sensors" tab to view the list of installed sensors.
4.
Scroll down to see the VOC sensor parameters.
5.
Click on "Edit"
6.
Click 3-Point Calibration (the check mark should now be showing).
7.
Click the
8.
When you are done, quit Safety Suite Device Configurator (SSDC) and then press [Y/+]
on the MultiRAE to exit the PC communications mode. The instrument returns to
operating in Normal mode.
Honeywell MultiRAE Series™
to set sensor parameters.
icon.
129
User Manual