Off/On (off by default)
l
Motionless Time: time the instrument is motionless before initiating a pre-alarm (30
l
seconds by default)
Motion Sensitivity: set to low, medium, or high to compensate for ambient vibration or
l
motion (medium by default)
Warning Time: countdown, in seconds, from pre-alarm to Man Down alarm (30 seconds
l
by default)
Falldown Time: time the instrument is falldown before initiating a pre-alarm (30 seconds
l
by default)
Falldown Sensitivity (low, medium, high, or custom)
l
Settings are adjustable in the Alarms menu on the MultiRAE but can also be changed and
saved by using Safety Suite Device Configurator (SSDC) software.
When the Man Down pre-alarm is activated, the buzzer sounds and LEDs flash twice per
second, and a countdown begins.
If the MultiRAE's user presses [Y/+] for "Yes" in response to the "Are You OK?" question
l
on the screen before the countdown reaches zero, the Man Down alarm stops and the
main reading screen is displayed.
If the person does not press [Y/+] for "Yes" in response to the "Are You OK?" question on
l
the screen before the countdown reaches zero, the Man Down alarm is triggered.
If the person presses [N/-] during the countdown, answering the "Are You OK?" question
l
with "No," the Man Down alarm starts.
If wireless connectivity is enabled, a Man Down message is also sent to remote observers.
IMPORTANT!
When gas or radiation alarm conditions exist at the same time as the Man Down is activated,
the pre-alarm stage is skipped and the instrument goes straight into a Super Alarm (gas or
radiation and Man Down) with four beeps/flashes per second.
IMPORTANT!
MultiRAE & MultiRAE Pro set up for benzene sampling: Man Down alarm capability is off during
Benzene Mode measurements.
IMPORTANT!