Honeywell MultiRAE Series User Manual page 136

Portable multi-gas monitor
Table of Contents

Advertisement

Single-Sensor Span Calibration
To perform span calibration of an individual sensor, follow these steps:
1.
At the Calibration Menu, select "Single Sensor Span."
2.
Select a sensor from the list.
3.
Connect the calibration adapter and connect it to a source of calibration gas.
4.
Verify that the displayed calibration value meets the concentration label on the gas
cylinder.
5.
Start the flow of calibration gas.
6.
Press [Y/+] to start calibrating. You can abort the calibration at any time during the
countdown by pressing [N/-].
Honeywell MultiRAE Series™
131
User Manual