Roaming
The Roaming function provides continuous wireless connectivity between zones enabling users
of wireless monitors to travel from one zone/work area to another without losing
communication between their monitor and Safety Suite Device Configurator (SSDC). You can
turn Roaming on and off in your monitor, or you can set it via SSDC.
1.
Press [N/-] to scroll down to "Roaming."
2.
Press [Y/+] to select "Roaming."
3.
Press [Y/+] to scroll to "On" or "Off."
4.
Press [Y/+] to "Save."
Note: When Roaming is on, you cannot change the instrument's PAN ID. When Roaming is on,
the "PAN ID" is not shown in the Wireless menu. To turn "PAN ID" back on:
1.
In the Wireless menu, press [N/-] to scroll down to "Roaming."
2.
Press [Y/+] to select "Roaming."
3.
Press [Y/+] to select "Off."
4.
Press [Y/+] to save the change.
Roaming is now off, and the "PAN ID" menu item is now visible and selected.
Notice: Roaming is not available for LoRa
PAN ID
The MultiRAE and any other devices that you want to interconnect wirelessly must have the
same PAN ID. You can set the PAN ID in the instrument or through Safety Suite Device
Configurator (SSDC). Note: When Roaming is turned on, the PAN ID menu item is unavailable
and therefore the PAN ID cannot be changed. To make the menu item available and to change
the PAN ID in the instrument, turn Roaming off.
1.
Press [Y/+] to increase the number and [N/-] to advance to the next digit.
2.
After moving to the last digit and making changes, press [MODE].