Honeywell MultiRAE Series User Manual page 65

Portable multi-gas monitor
Table of Contents

Advertisement

RAE Systems by Honeywell specifically disclaims liability for all loss or damage arising out of
any use of the RAE-Sep Tube Cartridge that violates the warnings and instructions in this
manual.
Separation Tube Cartridge Installation
Before using the MultiRAE or MultiRAE Pro monitor equipped with a 9.8eV lamp for benzene-
specific measurement, you must attach a RAE-Sep Tube Cartridge to the instrument. Follow
this procedure:
1.
Remove the dust filter from the instrument (if one is currently attached).
2.
Align a new RAE-Sep Tube Cartridge with the instrument's inlet.
3.
Screw on the RAE-Sep Tube Cartridge until it is seated snugly in place.
4.
The "B" switch of the RAE-Sep Tube Cartridge should be aligned with the triangle on the
inlet.
IMPORTANT!
Do not overtighten any portion of the sampling assembly.
Note: When the MultiRAE or MultiRAE Pro monitor equipped with a 9.8eV lamp is used only for
VOC monitoring, there is no need to use a RAE-Sep Tube Cartridge. Make sure the cartridge is
set for TVOC operation (the letter "T" should be visible).
Always rotate the RAE-Sep Tube Cartridge clockwise. Turning it counterclockwise loosens it
from the inlet.  Rotate the cartridge so that a new (unused) tube is aligned with the "B" marker.