Honeywell MultiRAE Series User Manual page 80

Portable multi-gas monitor
Table of Contents

Advertisement

Single Sensor Span
Instead of performing a span calibration on multiple sensors simultaneously, you can select a
single sensor and perform a span calibration.
Note: If a calibration icon (bottle with bottom portion filled in) is shown next to any of the
sensors, it means that the sensor is due for a full calibration.
To perform span calibration of an individual sensor, follow these steps:
1.
At the Calibration Menu, select "Single Sensor Span."
2.
Select a sensor to calibrate from the list.
3.
Install the calibration adapter and connect it to a source of calibration gas.
4.
Verify that the displayed calibration value meets the concentration specified on the gas
cylinder.
5.
Start the flow of calibration gas.
6.
Press [Y/+] to start calibrating or wait for calibration to start automatically.
7.
You can abort the calibration at any time during the countdown by pressing [N/-].
Note: Dotted line indicates automatic progression.
8.
If the calibration is not aborted, the display shows the sensor names and tells you
whether the calibration passed or failed, followed by the sensor readings.
Honeywell MultiRAE Series™
75
User Manual