Honeywell MultiRAE Series User Manual page 83

Portable multi-gas monitor
Table of Contents

Advertisement

Single Sensor Bump
This menu allows a bump test to be performed on an individual sensor of your choice.
Note: If a bump test icon (bottle with bottom portion not filled in) is shown next to any of the
sensors, it means that the sensor is due for a bump test.
To perform a bump test on an individual sensor, follow these steps:
1.
At the Calibration Menu, select "Single Sensor Bump."
2.
Scroll down the list using [N/-], and then press [Y/+] to select a sensor to bump.
3.
Install the calibration adapter and connect it to a source of calibration gas.
4.
Verify that the displayed calibration value meets the concentration specified on the gas
cylinder.
Important!
After each sensor is bump tested, and you press "OK," the next sensor in the menu list is
highlighted.
Note: Dotted line indicates automatic progression.
5.
Press [Y/+] to start bump or wait for bump test to start automatically.
6.
Start the flow of calibration gas.
Honeywell MultiRAE Series™
78
User Manual