Honeywell MultiRAE Series User Manual page 89

Portable multi-gas monitor
Table of Contents

Advertisement

Change Meas. Gas
The MultiRAE has extensive onboard gas libraries for combustible gases and VOCs that you
can use to configure your MultiRAE to automatically apply the appropriate correction factors
and produce readings in the units of the desired combustible gas or VOC.
Measurement gases are organized in four lists*:
My List is a customized list of gases that you create. It contains a maximum of 10 gases
l
and can only be built in Safety Suite Device Configurator (SSDC) on a PC and
transferred to the instrument.
Note: The first gas in the list is always isobutylene (it cannot be removed from the list).
Last Ten is a list of the last ten gases used by your instrument. The list is built
l
automatically and is only updated if the gas selected from Custom Gases or Library is
not already in the Last Ten. This ensures that there is no repetition.
Gas Library is a library that consists of more than 200 gases for the PID sensor and
l
more than 50 for the catalytic LEL sensor.
Custom Gases are gases with user-modified parameters. Using Safety Suite Device
l
Configurator (SSDC), all parameters defining a gas can be modified, including the name,
span value(s), correction factor, and default alarm limits.
* See RAE Systems by Honeywell Technical Note TN 156 (for LEL) and TN 106-B (for PID) for
more details.
Honeywell MultiRAE Series™