7.
Gently pry up the cap using a rocking motion:
8.
Once the cap is removed, set it aside.
9.
Now lift the sensor electrode panel from the module:
10.
Clean the sensor electrode panel (crystal) in a solution of isopropanol or methanol lamp
cleaner (included, along with cleaning swabs, in a PID Lamp Cleaning Kit (PN: 081-
0017-000), sold separately), and allow it to dry.
11.
Clean the lamp's window with a cleaning swab dipped in isopropanol or methanol lamp
cleaner and allow it to dry. Do not touch the lamp window with your fingers, as the
residual oils will affect its performance and shorten its life.