or deactivated from this screen.
Must Calibrate. The user is prompted to calibrate the instrument when calibration is due (as set
by the calibration interval). There are two programmable options:
Can't Bypass. Unless calibration is performed, the instrument cannot be used, and the
l
only option is to turn off the instrument.
Can Bypass. If calibration is due but the user does not want to perform a calibration, the
l
instrument can still be used. In this case, the instrument records that the user has
bypassed the calibration requirement in a Policy Violation report.
Must Bump. The user is prompted to bump test the instrument when a bump test is due (as set
by the bump test interval). There are two programmable options:
Can't Bypass. Unless a bump test is performed, the instrument cannot be used, and the
l
only option is to turn off the instrument.
Can Bypass. If a bump test is due but the user does not want to perform one, the
l
instrument can still be used. In this case, the instrument records that the user has
bypassed the bump testing requirement in a Policy Violation report.
These are the screens that are shown on a MultiRAE after startup if "Can Bypass" is selected:
If "Can't Bypass" is selected, the display looks like this, and only allows the options of
performing the test or shutting down: