Honeywell MultiRAE Series User Manual page 121

Portable multi-gas monitor
Table of Contents

Advertisement

or deactivated from this screen.
Must Calibrate. The user is prompted to calibrate the instrument when calibration is due (as set
by the calibration interval). There are two programmable options:
Can't Bypass. Unless calibration is performed, the instrument cannot be used, and the
l
only option is to turn off the instrument.
Can Bypass. If calibration is due but the user does not want to perform a calibration, the
l
instrument can still be used. In this case, the instrument records that the user has
bypassed the calibration requirement in a Policy Violation report.
Must Bump. The user is prompted to bump test the instrument when a bump test is due (as set
by the bump test interval). There are two programmable options:
Can't Bypass. Unless a bump test is performed, the instrument cannot be used, and the
l
only option is to turn off the instrument.
Can Bypass. If a bump test is due but the user does not want to perform one, the
l
instrument can still be used. In this case, the instrument records that the user has
bypassed the bump testing requirement in a Policy Violation report.
These are the screens that are shown on a MultiRAE after startup if "Can Bypass" is selected:
If "Can't Bypass" is selected, the display looks like this, and only allows the options of
performing the test or shutting down:
Honeywell MultiRAE Series™
116
User Manual

Advertisement

Table of Contents
loading

Table of Contents