Honeywell EK280 Operating Instructions Manual page 58

Volume conversion device
Hide thumbs Also See for EK280:
Table of Contents

Advertisement

58
5.3.1.15 Deleting the measurement archive
Deleting the archive is not absolutely necessary when putting
the device into operation.
Move the cursor to the "Serv." register and to the "Clr.A" value (delete
measurement archive) via the following path:
Serv.
All measurement archives (no logs) will be deleted.
In order to ensure that the archive is not accidentally deleted,
the serial number of the EK280 must be entered whilst the
calibration lock is open (the number is located on EK280
identification plate).
Press the ENTER button.  The value will start to flash.
Move the cursor using the arrow keys  or  to the digits and enter
the serial number of the device using the arrow keys  or .
After you have entered the serial number of the device, press the
ENTER button to confirm the input. The input value can be deleted by
pressing the ESC button.
5.3.1.16 Deleting the certification data log
The log can only be deleted if the calibration lock is open.
Move the cursor to the "Serv." register and to the value "ClCDL"
(delete certification data log) via the following path:
Serv.  Change and Delete  ClCDL
Press the ENTER button.  "0" will start to flash.
Change the value to "1" with the arrow keys  or .
After you have entered the value, press the ENTER button to confirm
the input. The input value can be deleted by pressing the ESC button.
5.3.1.17 Closing and securing the calibration lock
After all settings subject to calibration regulations have been adjusted,
close the calibration lock in the same way that it was opened: Press the
buttons described in chapter 5.3.1.1 once again. The calibration switch is
secured with an adhesive label as described in chapter 5.3.2.2.
Assembly, Connection and Putting into Operation
Edit and delete
Clr.A

Advertisement

Table of Contents
loading

Table of Contents