52
5.3.1.6
Setting the volume at measurement conditions
In order to control the recording of the volume at measurement conditions,
the EK280 meter can be set once to the same value as the gas meter if the
administrator lock is open. It is always possible to set the volume if the
calibration lock is open:
–
Open the administrator lock or the calibration lock
–
Move the cursor to the „Serv." tab and go to the value „VmA"
(adjustable volume at measurement conditions) via the following
path:
Serv.
Press the ENTER key. One of the value digits flashes.
–
–
Move the cursor using the arrow keys or to the digits and
change these using the arrow keys or .
–
Once you have changed all of the digits, press the ENTER key in
order to confirm your entry. The entry can be cancelled by pressing
the ESC key.
–
In order to accept the set value, move the cursor to the value „Store"
Press the ENTER key. „0" flashes.
–
Change the value using the arrow keys or to „1"
–
–
Once you have changed the value, press the ENTER key in order to
confirm your entry. The entry can be cancelled by pressing the ESC
key. „Vm" is assumed for „VmA".
If, when the administrator lock is open, the assumption of
„VmA" for „Vm" is rejected with the message –13--, then this
process has already been carried out once.
Other changes are possible if the calibration lock is open or, if
the administrator lock is open, after deleting the change
information for comparison of „Vm" with the gas meter (see
chapter 5.3.1.7).