GE AKTA avant Operating Instructions Manual page 64

Preparative chromatography system
Hide thumbs Also See for AKTA avant:
Table of Contents

Advertisement

4 Installation
4.1 Site preparation
4.1.6 Required materials
Fraction collector tubes and
bottles
The tubes and bottles used in the built-in fraction collector must fulfill the requirements
listed in the following table. Examples of manufacturers are also listed in the table.
Tube or bottle
size (ml)
3
5
8
15
50
250 ml bottle
1
Deep well plates, requirements
The deep well plates used in the built-in fraction collector must fulfill the requirements
listed in the table below.
Property
No. of wells
Shape of wells
Well volume
64
Diameter (mm)
Min.
10.5
10.5
12
16
28
L: 55
1
W: 55
Length and width of the rectangular bottle base
Height (mm)
Max.
Min.
11.5
50
12
70
13.3
96
17
114
30
110
L: 64.5
-
1
W: 64
Specification
24, 48, or 96
Square, not cylindrical
10, 5, or 2 ml
ÄKTA avant Operating Instructions 29101556 AB
Examples of manu-
facturers
Max.
56
NUNC™
76
VWR™
102
BD™ Biosciences,
VWR
120
BD Biosciences
116
BD Biosciences
121
Nalgene™, Kautex™