3 System description
3.1 ÄKTA avant instrument overview
Module
Sample pump P9-S or P9H S
Pressure monitor R9
Pump flow restrictor
Mixer M9
Inlet valve A V9-IA or V9H-IA
Inlet valve B V9-IB or V9H-IB
Quaternary valve Q9
Sample inlet valve V9-IS or
V9H-IS
Injection valve V9-Inj or V9H-Inj
42
Description
A high precision pump which delivers sample or
buffer in purification runs.
Pressure monitor which reads the system pressure
after System Pump A and System Pump B.
Prevents the system from siphoning when the flow
path after the pump is open. Gives a small back
pressure to the pump in extreme low pressure
applications.
Mixes the buffers delivered from the system
pumps to a homogeneous buffer composition.
Three mixer chambers are available for ÄKTA
avant 25. Available volumes are: 0.6 ml, 1.4 ml
(mounted at delivery) and 5 ml.
Three mixer chambers are available for ÄKTA
avant 150. Available volumes are: 1.4 ml, 5 ml
(mounted at delivery), and 15 ml.
CAUTION
Risk of explosion. Do not
use Mixer chamber 15 ml
with an ÄKTA avant 25 sys-
tem configuration. The
maximum pressure for Mix-
er chamber 15 ml is 5 MPa
(50 bar).
Inlet valve for System Pump A with seven inlet
ports and integrated air sensor.
Inlet valve for System Pump B with seven inlet
ports and integrated air sensor.
Valve which allows automatic mixing of four dif-
ferent solutions.
Inlet valve for sample solution, with eight inlet
ports (seven sample inlets and one buffer inlet)
and integrated air sensor.
Valve which directs sample onto the column.
ÄKTA avant Operating Instructions 29101556 AB