Power failure
The result of a power failure depends on which unit is affected.
Power failure to...
ÄKTA avant instrument
Computer
ÄKTA avant Operating Instructions 29101556 AB
then...
Push the Power switch to the O position,
or
disconnect the power cord from the wall socket.
Result: The run is interrupted immediately.
Note:
The sample and data may be lost as a result of switching off the
power.
will result in...
The run is interrupted immediately
Data collected up to approx. 5 seconds before
the power failure is available in UNICORN.
The UNICORN computer shuts down
The instrument display shows status Not
connected
The run is interrupted immediately
Data generated up to 10 seconds before the
power failure can be recovered
Note:
The UNICORN client may lose connection to the
instrument during a temporary overload of the
processor and display an error message. This
may appear as a computer failure. The run
continues and you can restart the UNICORN
client to regain control. No data will be lost.